Value collected
Measurement taken at this point.
See Also

Indicator state
The name assigned to the indicator state, for example, “Normal” or “Critical”.
See Also

Alarm Type
The type of warning or alarm that is triggered when an indicator reading falls within the range for this state.
See Also

Fixed
Sometimes a problem identified by an indicator reading is fixed during the inspection when the reading is collected. In this case, the reading does not raise an alarm in the system because no further action is required.
See Also

Reading status
The status of the reading can be:
•
Unread
•
Read
•
Processed
See Also

Technology
Technology used to take the measurement point reading.
See Also

Measurement device
If the indicator uses a technology associated with measurement device assets, you can select the measurement device used to take the reading.
See Also

Severity
Value assigned by an external inspector to assist an in-house representative, such as an engineer, to make the final assessment of the findings. Values might include “Slight”, “Moderate”, “High”, and “Excessive”.
See Also

Recommended action
Value assigned by an external inspector to assist an in-house representative, such as an engineer, to make the final assessment of the findings. Values might include “No further action required” and “Requires immediate intervention”.
See Also

Notes
Information provided by the collector of the reading.
See Also

Alarm name
The name of the indicator alarm. You can enter up to 50 characters to describe the alarm.
See Also

Note
The comments entered by the employee who acknowledged the reading.
See Also

Fixed
Sometimes a problem identified by an indicator reading is fixed during the inspection when the reading is collected. In this case, the reading does not raise an alarm in the system because no further action is required.
See Also

Image 2
You can attach up to five image files, for example, photographs, to each measurement point reading.
See Also

Image 3
You can attach up to five image files, for example, photographs, to each measurement point reading.
See Also

Image 4
You can attach up to five image files, for example, photographs, to each measurement point reading.
See Also

Image 5
You can attach up to five image files, for example, photographs, to each measurement point reading.
See Also